Difference between revisions of "Zeta Potential Analyzers"

Jump to navigation Jump to search
3,480 bytes removed ,  06:35, 18 January 2013
no edit summary
Line 1: Line 1:
[[Category:Analyzing]]{{Knoppen}}
[[Category:Analyzing]]{{Knoppen}}
[[File:Zeta potential analyzer.JPG|thumb|right|Zeta Potential Analyzers]]
'''Zeta Potential Analyzers''' is an analytical device for characterizing zeta potential as a property of interfacial layer in liquid particulates (dispersions, emulsions) and porous bodies. There is ISO Standard -13099 being prepared that would describe the main principles underlying functioning of such devices. Zeta potential is not directly measurable parameter.  It should be calculated from other experimentally measurable properties of the system using appropriate theory. There is no zeta potential analyzer that would be capable of working with all possible liquid based heterogeneous systems. All of them have limitation. It might be quite challenging to find an optimum device for particular application.


Zeta potential is a scientific term for electrokinetic potential[1] in colloidal systems. In the colloidal chemistry literature, it is usually denoted using the Greek letter zeta, hence ζ-potential. From a theoretical viewpoint, zeta potential is electric potential in the interfacialdouble layer (DL) at the location of the slipping plane versus a point in the bulk fluid away from the interface. In other words, zeta potential is the potential difference between the dispersion medium and the stationary layer of fluid attached to the dispersed particle.
A value of 25 mV (positive or negative) can be taken as the arbitrary value that separates low-charged surfaces from highly-charged surfaces.
The significance of zeta potential is that its value can be related to the stability of colloidal dispersions (e.g., a multivitamin syrup). The zeta potential indicates the degree of repulsion between adjacent, similarly charged particles (the vitamins) in a dispersion. For molecules and particles that are small enough, a high zeta potential will confer stability, i.e., the solution or dispersion will resist aggregation. When the potential is low, attraction exceeds repulsion and the dispersion will break and flocculate. So, colloids with high zeta potential (negative or positive) are electrically stabilized while colloids with low zeta potentials tend to coagulate or flocculate as outlined in the table.[2][3]


==Application Fields==
* Environment:Water treatment, waterworks, flocculation agents, dispersants, flocculation control, microorganisms, planktons, asbestos, bubbles, soil, beneficiation technology, and etc.
* Nano-materials:Functional material development,Arecording materials, pigments, ceramics, catalytic agents, polymers, carbon black, carbon nanotubes, inkjet inks, aqueous/non-aqueous solvent paints, fuel cells, coating materials (for automobile parts / electronic parts), paper manufacturing, surface active agents, and etc.
* Life Science:Erythrocyte, biological cells, protein, DDS, liposome, vesicle, drug discovery, and etc.


Measurement of zeta potential
 
Zeta potential is not measurable directly but it can be calculated using theoretical models and an experimentally-determined electrophoretic mobility or dynamic electrophoretic mobility.
 
Electrokinetic phenomena and electroacoustic phenomena are the usual sources of data for calculation of zeta potential.
 
[edit]Electrokinetic phenomena
==Features==
Main article: Electrokinetic phenomena
* Zeta potential measurement
Electrophoresis is used for estimating zeta potential of particulates, whereas streaming potential/current is used for porous bodies and flat surfaces. In practice, the Zeta potential of dispersion is measured by applying an electric field across the dispersion. Particles within the dispersion with a zeta potential will migrate toward the electrode of opposite charge with a velocity proportional to the magnitude of the zeta potential.
:* Microscopic electrophoresis method
This velocity is measured using the technique of the Laser Doppler Anemometer. The frequency shift or phase shift of an incident laser beam caused by these moving particles is measured as the particle mobility, and this mobility is converted to the zeta potential by inputting the dispersant viscosity and dielectric permittivity, and the application of the Smoluchowski theories (see below).[5]
:* Automatic tracking in image processing
[edit]Electrophoresis
:* Visual and manual measurement of individual particles
Main article: Electrophoresis
* Image capture of particles
Electrophoretic velocity is proportional to electrophoretic mobility, which is the measurable parameter. There are several theories that link electrophoretic mobility with zeta potential. They are briefly described in the article on electrophoresis and in details in many books on colloid and interface science.[6][7][8][9] There is an IUPAC Technical Report[10] prepared by a group of world experts on the electrokinetic phenomena.
:* Image archive or remeasurement function
From the instrumental viewpoint, there are two different experimental techniques:
* PC-controlled instrument
Microelectrophoresis. It has the advantage of yielding an image of the moving particles. On the other hand, it is complicated by electro-osmosis at the walls of the sample cell.
:* Stage control or voltage feedback control or switching control for optical system
Electrophoretic light scattering. It is based on dynamic light scattering. It allows measurement in an open cell which eliminates the problem of electro-osmotic flow for the case of an Uzgiris, but not a capillary cell. And, it can be used to characterize very small particles, but at the price of the lost ability to display images of moving particles.
Both these measuring techniques may require dilution of the sample. Sometimes this dilution might affect properties of the sample and change zeta potential. There is only one justified way to perform this dilution - by using equilibrium supernatant. In this case the interfacial equilibrium between the surface and the bulk liquid would be maintained and zeta potential would be the same for all volume fractions of particles in the suspension. When the diluent is known (as is the case for a chemical formulation), additional diluent can be prepared. If the diluent is unknown, equilibrium supernatant is readily obtained by centrifugation.
[edit]Electroacoustic phenomena
Main article: Electroacoustic phenomena
There are two electroacoustic effects that are widely used for characterizing zeta potential: colloid vibration current and electric sonic amplitude, see reference.[8] There are commercially available instruments that exploit these effects for measuring dynamic electrophoretic mobility, which depends on zeta potential.
Electroacoustic techniques have the advantage of being able to perform measurements in intact samples, without dilution. Published and well-verified theories allow such measurements at volume fractions up to 50%, see reference. Calculation of zeta potential from the dynamic electrophoretic mobility requires information on the densities for particles and liquid. In addition, for larger particles exceeding roughly 300 nm in size information on the particle size required as well.